KM KIT
Contamination Test Unit CTU 1000 Series
CTU 10xx, CTU 12xx
Order Code Example: CTU-1040-M-Z-Z, CTU-1240-M-Z-Z
HYDAC

Contamination Test Unit CTU 1000 Series CTU 10xx, CTU 12xx Order Code Example: CTU-1040-M-Z-Z, CTU-1240-M-Z-Z HYDAC

Universal equipment

Price on request

Request pricing and availability

Contact our specialists for details, compatibility and supply options.

Description

1. Function: Determines the technical cleanliness of lightly contaminated components via wet extraction (spray, ultrasound) and filtration onto a membrane. 2. Measuring Range: Analysis of contamination mass and particle size. Blank values depend on environment (0.1-0.8 mg). 3. Accuracy: Dependent on procedure and environment. 4. Process Connection: Component placed in analysis chamber (clean box). Test liquid sprayed or used with ultrasound. 5. Output & Supply: PC controlled with software. Automatic filtration. Requires external compressed air (6.5-7.0 bar) and power supply. 6. Electrical Connection: According to order (various regional voltages). 7. Certifications: Not explicitly stated. 8. Housing: Mobile unit on casters with analysis chamber and storage/filtration module. 9. Materials: Analysis chamber: polished stainless steel. Fine filtration to ISO 12/9 (1 µm). 10. Options: Size (10xx/12xx), Version (2011/2014), Test liquid (solvent/water), Extraction method (spray Z / ultrasound U), Supplementary details (external rinsing, fluid connections, etc.).

Specifications

CategoryUniversal equipment
Product ID6711111011-1738
ManufacturerHYDAC
← Back to Catalog